Fujikura Optimized sensors for non-invasive blood pressure monitoring |
|
|
|
|
MODEL SELECTION
XFPM (Gauge) Package
(8 mm tube length)
Best used with silicon tubing or directly attached to manifold
XFGM-3 (Gauge) Package
(3 mm tube length)
Best used directly attached to manifold and sealed with an O Ring
FEATURES
- Pressure ranges and performance optimized to meet regulatory standards
- Accurate, reliable, competitively priced
- Small package sizes
- High level output signal saves space and adds reliability
- Operational amplifiers and temperature compensation thin-film-resistors are integrated on chip.
- Offset voltage and sensitivity are pre-calibrated.
SPECIFICATIONS
- Rated Pressure: 0 to 300 or 370 mmHg
- Cuff pressure measurement
- Part Numbers: XFPM-050KPG(R)-P1, XFGM-3050KPGSR-P1, XFGM-6050KPGSR-P1
- 0 to 300 mmHg
- Accuracy: +/- 3 mmHg
- Requires offset and auto-zero calibration
- Part Numbers: XFPM-050KPG(R)-P3, XFGM-3050KPGSR-P3, XFGM-6050KPGSR-P3
- 0 to 300 mmHg
- Accuracy: +/- 2.0 mmHg @ 150 mmHg and +/- 2.5 mmHg @ 300 mmHg
- Requires offset and span and auto-zero calibration
- Overpressure measurement
- Part Numbers: XFPM-050KPG(R)-BP, XFGM-3050KPGSR-BP, XFGM-6050KPGSR-BP
- 0 to 370 mmHg
- Accuracy: +/- 5.5 mmHg
- Requires no calibration (unless higher accuracy is required)
Call Servoflo for detailed data sheets and application support: 781 862 9572
|